MESH NOMINAL ·
Equipment · Controlled Procedure

VC-SOP-0502 — PhytoSense Mesh Calibration

The procedure for verifying and calibrating the PhytoSense Mesh — the distributed sensor network measuring sap flow, turgor, VOC, and chlorophyll fluorescence — so that every downstream control loop and alarm acts on trustworthy data.

SOP ID: VC-SOP-0502 Rev: E Effective: 2091-05-08 Next Review: 2092-05-08 Owner: Instrumentation Engineer (OPS-ENG) Approver: Dr. P. L. Isley, CBO Class: Equipment · Metrology
Uncalibrated data drives bad control. Climate, irrigation, and containment alarms all consume PhytoSense readings. A drifted node can mask a real hazard or trigger false response. A node past its calibration window is flagged suspect and excluded from auto-control until verified.

1. Purpose

To define the controlled method for routine verification, calibration, and drift management of PhytoSense Mesh nodes so that measurement accuracy stays within tolerance, suspect nodes are isolated from control loops, and the calibration state of the network is fully traceable.

2. Scope

Applies to all PhytoSense Mesh node types across the facility: sap-flow, turgor-pressure, VOC, and fluorescence sensors, plus their MDF gateways. The data these feed governs VC-SOP-0102 (climate), VC-SOP-0118 (dosing), and VC-SOP-0402 (pressure cross-checks). Excludes the Spectral Canopy Array imaging chain (VC-SOP-0124).

3. Definitions

TermDefinition
NodeA single PhytoSense sensor unit reporting to an MDF gateway.
Reference StandardA traceable calibration source (gas mix, pressure standard, fluorescence target) of known value.
DriftSlow divergence of a node's reading from true value between calibrations.
Suspect NodeA node out of tolerance or past its window; excluded from auto-control pending action.
Calibration WindowThe maximum interval between calibrations for a given node type.
As-Found / As-LeftThe node reading before and after calibration; the pair documents drift.

4. Responsibilities

RoleResponsibility
Instrumentation Engineer (Owner)Owns the calibration program; approves tolerances and standards; signs out-of-tolerance dispositions; owns this SOP.
Calibration TechnicianPerforms verification and calibration; records as-found/as-left.
Watch CommanderAuthorizes removing a node from auto-control during calibration.
CHLORATracks calibration windows; flags suspect nodes; excludes them from control; logs results.

5. Equipment & Standards

6. Calibration Windows & Tolerances

Node TypeWindowToleranceAs-Found Fail Action
Sap-flow90 days±3 % of readingRecalibrate; review affected dosing logs
Turgor-pressure90 days±0.02 MPaRecalibrate; flag stress-alarm history
VOC60 days±5 % or ±2 ppbRecalibrate; cross-check neighbor nodes
Fluorescence120 days±2 % Fv/FmRecalibrate; re-baseline zone canopy
MDF gateway clock30 days±50 msResync; audit timestamp continuity

7. Procedure

7.1 Scheduling & Suspect Isolation

  1. CHLORA lists nodes due within the next 7 days; the Technician sequences work to minimize control-loop gaps.
  2. Before touching a node, the Watch Commander authorizes its removal from auto-control; CHLORA substitutes the nearest healthy node or holds the loop at last-good.
  3. Any node already flagged suspect is prioritized.

7.2 As-Found Verification

  1. Present the reference standard to the node without adjustment; record the as-found reading.
  2. Compare against tolerance (Section 6). If within tolerance, log as-found = as-left and proceed to release.
  3. If out of tolerance, record the deviation and notify the Instrumentation Engineer for disposition — including review of any control action taken on bad data.

7.3 Adjustment & As-Left

  1. Adjust the node against the reference standard across its operating range (minimum 3-point: low, mid, high).
  2. Record the as-left reading at each point; confirm all within tolerance.
  3. A node that cannot be brought into tolerance is tagged out-of-service and scheduled for replacement; CHLORA keeps it excluded.

7.4 Cross-Validation & Release

  1. Cross-check the calibrated node against at least one neighbor reading the same zone; agreement must be within combined tolerance.
  2. Verify the gateway timestamp is synced within ±50 ms before re-enabling control input.
  3. Watch Commander authorizes return to auto-control; CHLORA restores the node and stamps the new window.

7.5 Quality Checkpoints

8. Anomaly Responses

FindingIndicationImmediate Action
Window LapsedNode past calibration windowCHLORA flags suspect and excludes from control; schedule within 48 h.
Out of ToleranceAs-found exceeds toleranceDisposition with engineer; review control actions on the bad data window.
Neighbor DisagreementCalibrated node still disagrees with neighborsHold release; investigate sensor or local plant condition before trusting either node.
Gateway Time SkewTimestamp drift > 50 msResync; audit affected records for ordering errors; notify monitoring.

9. Records

10. References

11. Revision History

RevDateAuthorSummary
A–C2087–2089Instrumentation EngineerInitial issue; per-type windows and tolerances established.
D2090-09-17Instrumentation EngineerSuspect-node auto-exclusion and as-found/as-left record added.
E2091-05-08Instrumentation EngineerNeighbor cross-validation gate and gateway clock sync formalized. CBO approval.